Fresenius Medical Care 2008K Operator's Manual page 164

2008k hemodialysis machine operator's manual
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

164
Tx Clock Paused
Blood is sensed in optical detector
while Tx clock is paused
UF Goal greater than
Entry value for goal is out of range.
max. value
UF Goal Reached
This message is to alert the operator
that the preset ultrafiltration goal has
been reached.
UF Is On!
An action has been initiated that
requires the UF to be off.
UF Profile Error
A UF profile calculation error has
been detected
UF Pump Alarm
UF pump is not connected or is not
pulsing properly.
UF Rate Error
A calculation error has been detected
UF Removed cleared
This temporary message is displayed
when the Tx clock is turned on the
first time after the New Tx key was
pressed. The UF removed has been
set to 0.
UF Removed not cleared
This temporary message is displayed
when the Tx clock is turned on other
than the first time after the New Tx
key was pressed. The UF removed
has been not been set to 0.
Upper Dia. has been set
The operator has attempted to set the
to Min [Max]
upper diastolic pressure limit higher or
lower than allowed. The machine has
set the limit to the highest or lowest
value allowed.
Troubleshooting
Warning
Start Tx clock
Dialog
Readjust goal
Message
Warning
Reset the alarm. The UF Goal has been reached and the UF Rate
will drop to the minimum UF Rate. The operator must take further
action to comply with the prescribed treatment.
Advisory
To proceed with the selected operation, turn the UF pump off.
Alarm
Reset the UF parameters
Alarm
Take machine out of service and alert a qualified technician.
Alarm
Reset the UF parameters
Advisory
No action necessary.
Advisory
No action necessary.
Verify that the limit setting is acceptable
Dialog
Message
2008K Operator's Manual rev. K