Fresenius Medical Care 2008K Operator's Manual page 161

2008k hemodialysis machine operator's manual
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

Message
Purpose of Message
Set Blood Flow to 300
This message is a prompt for the
operator to set the blood flow rate in
preparation for the Access Flow test
Set TMP Limits?
This is a message to re-center the
TMP limits if necessary
Set Venous Limits?
This is a message to adjust the
venous limits if necessary
Short Power Down
The machine was switched off for 1 –
2 minutes and turned back on. Setup
values have not been set to default
values.
Single Needle On?
An action has been initiated that
requires the Single Needle option to
be off.
SN pump in use
The OLC test may not be run when
the Single Needle system is in use.
Standby For Test
This message is displayed before the
start of the Alarms and Pressure test
Start Na+ has been set to
The operator has attempted to set a
max.
Starting Na+ higher than allowed.
Start Na+ has been set to
The operator has attempted to set a
min.
Starting Na+ lower than allowed.
Super I/O no comm
Hardware related error message
Switch Bloodlines Back
This message is a prompt for the
operator return the bloodlines to their
original position
161
Type
Action Required
Warning
To proceed, set Blood Flow to 300 and press CONFIRM.
Advisory
Rising TMP may indicate a leak in the balancing system and should
be investigated.
Press and hold Override for 1 second to re-center the limits.
Advisory
Press and hold RESET for 1 second to adjust the limits.
Advisory
Verify that the dialysis parameters are as desired.
Warning
To proceed with the selected operation, de-select the Single Needle
option.
Advisory
Do not attempt an OLC test when using Single Needle
Advisory
Advisory only. No action is required.
Advisory
The Starting Na+ will be set to the highest allowed Na+ level. Check
that the value is acceptable.
Advisory
The Starting Na+ will be set to the lowest allowed Na+ level. Check
that the value is acceptable.
Opening
This will only affect the use of the Single Needle pump system. If
Screen
necessary, turn off the machine and try again. If the message is not
Message
cleared, alert a qualified technician.
Warning
To proceed, press CONFIRM.
Troubleshooting
2008K Operator's Manual rev. K