Fresenius Medical Care 2008K Operator's Manual page 143

2008k hemodialysis machine operator's manual
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

Message
Purpose of Message
E.02 (Arterial or SN
Flash ROM CRC error
Blood Pump Message)
E.03 (Arterial or SN
RAM check error
Blood Pump Message)
E.04 (Arterial or SN
Reference Voltage error
Blood Pump Message)
E.05 (Arterial or SN
EEPROM error
Blood Pump Message)
E.06 (Arterial or SN
Watchdog timeout
Blood Pump Message)
E.07 (Arterial or SN
+ 12 volts is outside the allowable
Blood Pump Message)
range
E.08 (Arterial or SN
+ 24 volts is outside the allowable
Blood Pump Message)
range
E.09 (Arterial or SN
- 12 volts is outside the allowable
Blood Pump Message)
range
E.10 (Arterial or SN
+ 5 volts is outside the allowable
Blood Pump Message)
range
50 ms second time period exceeded
E.14 (Arterial or SN
Blood Pump Message)
E.15 (Arterial or SN
Software task was not completed
Blood Pump Message)
correctly
E.23 (Arterial or SN
Pump rotor turning when it should not
Blood Pump Message)
be for a second time
E.97 (Arterial or SN
Error copying data into Flash ROM
Blood Pump Message)
while in service mode
143
Type
Action Required
Alarm
If this message occurs, take the blood pump module out of service
and alert a qualified technician.
Alarm
If this message occurs, take the blood pump module out of service
and alert a qualified technician.
Alarm
If this message occurs, take the blood pump module out of service
and alert a qualified technician.
Alarm
If this message occurs, take the blood pump module out of service
and alert a qualified technician.
Alarm
If this message occurs, take the blood pump module out of service
and alert a qualified technician.
Alarm
If this message occurs, take the blood pump module out of service
and alert a qualified technician.
Alarm
If this message occurs, take the blood pump module out of service
and alert a qualified technician.
Alarm
If this message occurs, take the blood pump module out of service
and alert a qualified technician.
Alarm
If this message occurs, take the blood pump module out of service
and alert a qualified technician.
Alarm
If this message occurs, take the blood pump module out of service
and alert a qualified technician.
Alarm
If this message occurs, take the blood pump module out of service
and alert a qualified technician.
Alarm
If this message occurs, take the blood pump module out of service
and alert a qualified technician.
Alarm
If this message occurs in dialysis mode, take the blood pump module
out of service and alert a qualified technician.
Troubleshooting
2008K Operator's Manual rev. K