Fresenius Medical Care 2008K Operator's Manual page 128

2008k hemodialysis machine operator's manual
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

128
Message
Purpose of Message
# of Tests has been set to
The operator has attempted to set the
min
number of OLC test lower than
allowed.
5 Minutes Flow Off
Dialysate flow has been off for five
minutes.
*** 5V HIGH ***
Electronic self-test, power supply
limits exceeded.
*** 5V LOW ***
Electronic self-test, power supply
limits exceeded.
10 Fill Pgm in 1 hr
Ten fill programs have occurred
during a one-hour period.
12V POWER FAIL
Electronic self-test, power supply
limits exceeded.
*** 24 V HIGH
Electronic self-test, power supply
limits exceeded.
*** 24V LOW
Electronic self-test, power supply
limits exceeded.
Troubleshooting
Type
Action Required
Dialog
The machine has set the limit to the lowest value allowed. Verify that
Message
the number of OLC tests is acceptable.
Warning
1) Press RESET to silence the alarm.
2) If you intend for the flow to be off, set the Dialysate Flow button
to Seq(uential). Otherwise, turn on the dialysate flow with the
On/Off key.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified technician.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified technician.
Warning
1) Check the dialysate lines, especially around the connectors and
dialysate filter, for air entering the system and correct the
problem.
2) Press RESET to clear the alarm. If unable to reset the alarm,
disconnect the dialysate lines and close the dialysate ports of the
dialyzer with clean caps. Return the blood to the patient and
replace the machine with another machine. Alert a qualified
technician.
Note: Using a conventional dialyzer at a high UF rate can
cause frequent Fill programs because of a high TMP.
Lowering the UF rate by decreasing the UF Goal may
solve the problem.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified technician.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified technician.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified technician.
2008K Operator's Manual rev. K