Preparing For Use
To degauss the display
To degauss the display
If the logic analyzer has been subjected to strong magnetic fields, the CRT
might become magnetized and display data might become distorted.
•
To correct this condition, degauss the CRT with a conventional
external television type degaussing coil.
To clean the logic analyzer
•
With the instrument turned off and unplugged, use mild soap and
water to clean the front and cabinet of the logic analyzer.
Harsh soap might damage the water-base paint. Do not immerse the logic
analyzer in water.
To test the logic analyzer
•
If you require a test to verify the specifications, the HP 1660C/CS/CP
Series Logic Analyzers Service Guide is required.
•
If you require a test to initially accept the operation, perform the
self-tests described in Troubleshooting in this chapter.
•
If the logic analyzer does not operate correctly, go to the flow charts
provided in Troubleshooting in this chapter.
13-6