Section II
Paragraphs 2-29 to 2-34
2-29. OPTION INSTALLATION.
2-30.
Wher
it is desired to add one or two of the
available
optional
features
to
a standard
4262A
instrument, perform the installation as follows:
Refer to option installation illustrations
on facing page.
a. Push LINE switch to off.
b. Remove instrument top cover.
c. Follow the appropriate paragraph below.
2-31. OPTION 001 BCD DATA OUTPUT
INSTALLATION.
a. Remove
the left side middle and lower blind
covers from the rear panel.
b. Install two 50-pin connector assemblies in the
openings.
c. Set
BCD
switch
of
SW1
on
A23
board
assembly
(RED/ORANGE
GUIDE,
P/N:
04262-66523
or 04262-66623) from OFF to
opposite position. This board is located third
from front on the right side.
d. Connect cable attached to A23 board (shown
below)
between
A23
and A35
BCD
Option
board assemblies (P/N:
04262-66535).
Install
A35 in RED/GREEN GUIDE
option
receptacle.
e. Plug 2 each
flat cable assemblies from A35
BCD
Option
board into connector boards of
rear panel connector assemblies.
f. Install instrument top cover.
Model 4262A
.2-32. OPTION 004 COMPARATOR
INSTALLATION.
Refer to Fig 2-4 for installation procedure.
2-33.
COUPLING OPTION 004 COMPARATOR
WITH OPTION 001 BCD DATA OUTPUT
INSTALLATION.
'
:
a.Set
CMP
(comparator)
and
BCD
option
switches
of SW1
ON
A23
board
assemblies
(RED/ORANGE
GUIDE,
P/N: 04262-66523
or
04262-66623)
from
OFF
to
opposite
position.
This
board
is located
third
from
front on the right side.
.Connect
cables
attached
to A23
board
be-
tween A23 and A24 comparator option BCD
board
assembly.
No
other
cable
assembly
change
is necessary
for this combination
of
options.
. Refer to Paragraphs 2-31 and 2-32 for other
installation procedures.
-
2-34. OPTION 101 HP-IB REMOTE CONTROL
AND DATA OUTPUT INSTALLATION.
. Remove
right
side
blind
covers
from
rear
panel.
.Install
connector
board
assembly
(P/N:
04262-66503) in the opening and mount with
washers and nuts included with assembly.
. Set the HP-IB switch of SW1
on A23 board
assembly from OFF to opposite position. The
A23
board is located on the right side third
from front.
.Connect
cable
assembly
attached
to
A25
board
between
A23
and
A25 HP-IB
option
board assemblies (P/N: 04262-66525). Install
A25
in
RED/GREEN
GUIDE
option
receptacle.
. Plug
flat
cable
assembly
from
connector
board assembly P/N: 04262-66503 into A25
board
assembly
(installed
in
RED/GREEN
GUIDE receptacle).
OPTION
101 IS NOT COMPATIBLE
WITH
OPTIONS
001 AND 004.
26