6.1.2 Configure data logger

1. Connect to your data logger by using Device Configuration Utility.
2. On the Datalogger tab, change the data logger PakBus Address and PakBus/TCP Password
to match the values entered in the Konect PakBus Router setup. The PakBus/TCP Password
will make the data logger authenticate any incoming or outgoing PakBus/TCP connection.
3. On the Network Services tab in the PakBus/TCP Client field, enter the DNS address and
Port number noted during the Konect PakBus Router setup.
4. On the PPP tab, set Config/Port Used to CS I/O SDC8 or RS-232, depending on how you
are connected to the data logger.
5. (Optional) On the PPP tab, set User Name and Password if required by your cellular carrier
(usually outside of the United Sates).
6. Verify the Modem Dial String setting is blank.
7. If connecting through RS-232, on the Comport Settings tab, set RS232 BaudRate to
115200 Fixed.
8. Click Apply to save the changes. Verify the settings in the summary window.
(Recommended) Save a copy of the settings to a file on the computer. Click OK.
9. Shut down Device Configuration Utility and start it again. This will activate the Cellular tab
needed for the next step.
CELL200-Series 4G LTE Cellular
Module     12