Test 1: Bpm (Beats Per Minute) - Nellcor OxiMax SRC-MAX Technical Manual

Pulse oximetry functional tester
Table of Contents

Advertisement

SRC-MAX Operation with a Pulse Oximeter or Monitor
Pulse/Beats per Minute
Light Level
% Sp0
Percent Modulation
8. The SRC-MAX is ready to test the pulse oximeter or

Test 1: BPM (Beats per minute)

The BPM test checks the pulse oximeter or monitor beats per
minute operation at 60 and 200 BPM.
Prerequisite Procedure — Test Setup on page 12.
1. The pulse oximeter or monitor should be displaying a
2. Press and release the SRC-MAX Heart Rate/BPM button.
3. The BPM LED indicator on the SRC-MAX will switch to
4. Observe the display of the pulse oximeter or monitor.
14
2
monitor.
BPM of 60 (57 to 63 BPM).
illuminate the 200 numeric reading.
The numeric reading for BPM will gradually rise to the
200 (197 to 203 BPM) level. Allow the pulse oximeter or
monitor 10-30 seconds to achieve the test result.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents