Peripheral Devices
Millimeter-Wave Source Modules
Configuring the Signal Generator
The following procedure configures a PSG for use with any external source module that has a WR
(waveguide rectangular) frequency range of 90- 140 GHz. You can modify the frequency range to
match your source module.
1. Turn on the signal generator's line power.
NOTE
Automatic leveling at the source module output is not available with the OEM Source
Module selection.
2. Press
>
Frequency
(3 of 3)
3. Toggle the
OEM Source Off On
4. Press
OEM Source Module Config
The selections in the Standard WR Freq Bands menu are pre- defined frequency ranges and
multipliers for the most common external source module frequency ranges. They are provided for
setup convenience. If your source module has a frequency range not listed in the list of
pre- defined setups, use the
range and PSG frequency display multiplier. For more information on these manual settings, refer
to the E8257D/67D PSG Signal Generators Key Reference.
5. If the RF OFF annunciator is displayed, press
To obtain flatness- corrected power, refer to
page
133.
278
>
. Toggle the
Source Module
softkey to On.
>
Standard WR Freq Bands
,
Min Band Freq
Max Band Freq
"Creating and Applying User Flatness Correction" on
Agilent 8355x Source Module Off On
>
WR8 90-140GHz.
, and
Freq Multiplier
.
RF On/Off
softkey to Off.
softkeys to manually set the
Chapter 12