7. Press
>
# of Points
10
Steps 4, 5, and 6 enter the desired flatness–corrected frequencies into the step array.
8. Press
>
Return
Load Cal Array From Step Array
This populates the user flatness correction array with the frequency settings defined in the step
array.
9. Press
>
Amplitude
More (1 of 2)
Enter a nominal (average) value for the gain or loss of any cables or other devices connected
between the signal generator's RF output and the power sensor. Refer to figure
is entered as a positive number while loss is entered as a negative number.
10. Press
.
RF On/Off
This activates the RF output and the RF ON annunciator is displayed on the signal generator.
Perform the User Flatness Correction
NOTE
If you are not using an Agilent E4416A/17A/18B/19B power meter, or if your power meter
does not have a GPIB interface, you can perform the user flatness correction manually. For
instructions, see
1. Press
More (1 of 2)
>
Pressing the
Do Cal
routine. A progress bar is shown on the front panel display as the routine runs. The routine
generates a table of correction points with each correction point consisting of a frequency and
amplitude correction value for that frequency. The correction value at each point is the difference
between the power level, as measured by the power meter, and the output power of the signal
generator.
NOTE
A power meter timeout may occur at low power levels. If a power meter timeout error
message appears, increase the timeout value by pressing
More (1 of 2) > Meter Timeout
2. Press
.
Done
Pressing the
softkey loads the amplitude correction values into the user flatness correction
Done
array.
If desired, press
Configure Cal Array
This opens the user flatness correction array, where you can view the stored amplitude correction
values. The user flatness correction array title displays User Flatness: (UNSTORED) indicating
that the current user flatness correction array data has not been saved to the memory catalog.
Performing the User Flatness Correction Manually
If you are not using an Agilent E4416A/17A/18B/19B power meter, or if your power meter does not
have a GPIB interface, complete the steps in this section and then continue with the user flatness
correction tutorial.
Chapter 4
>
.
Enter
>
>
.
Ampl Offset
"Performing the User Flatness Correction Manually" on page
User Flatness
>
Do Cal
.
softkey causes the signal generator to perform the user flatness correction
.
.
Confirm Load From Step Data
Amplitude
Optimizing Performance
Creating and Applying User Flatness Correction
.
Figure 4-
137.
>
More (1 of 2)
>
4. Gain
User Flatness >
137