Page 2
Notice The material in this document is provided “as is,” and is subject to change without notice in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied with regard to this manual and to any of the Agilent products to which it pertains, including but not limited to the implied warranties of merchantability and fitness for a particular purpose.
Safety Information Instrument Markings The following markings are used on the signal generator. Familiarize yourself with each marking and its meaning before operating the signal generator. The instruction manual symbol. The product is marked with this symbol when it is necessary for the user to refer to the instructions in the manual.
Safety Information Warnings, Cautions, and Notes Warnings, Cautions, and Notes The following safety notations are used throughout this manual. Familiarize yourself with each notation and its meaning before operating the signal generator. WARNING Warning denotes a hazard. It calls attention to a condition or situation that could result in personal injury or loss of life.
Getting Started Checking the Shipment Checking the Shipment 1. Inspect the shipping container for damage. Signs of damage may include a dented or torn shipping container or cushioning material that shows signs of unusual stress or compacting. 2. Carefully remove the contents from the shipping container and verify that your order is complete. The following items are shipped standard with each signal generator: •...
Getting Started Meeting Electrical and Environmental Requirements Meeting Electrical and Environmental Requirements Environment The signal generator is designed for use in the following environmental conditions: • indoor use • < 15,000 feet (4,572 meters) altitude • 0 to 55° C temperature, unless otherwise specified •...
100/115 volts: 50/60/400 Hz nominal for 230/240 volts: 50/60 Hz nominal Power: E8257D PSG analog models: 300 watts maximum E8267D PSG vector models: 600 watts maximum CAUTION Damage to the signal generator may result if the supply voltage is not within the specified range.
Getting Started Meeting Electrical and Environmental Requirements AC Power Cord Localization The ac power cord included with the signal generator is appropriate to its geographic location. However, you can order additional ac power cords for use in different areas. The following table lists the available ac power cords, illustrates plug configurations, and identifies the geographic area in which each cord is appropriate.
Getting Started Configuring the Display Configuring the Display You can adjust the LCD display using features such as contrast, brightness, screen saver mode, and the screen saver delay. You can also toggle features such as inverse video, display updating in remote mode, and the screen saver on or off.
Getting Started Configuring the Display input to the front panel. The display light turns on again when any front panel key is pressed or when a remote command is sent. The screen saver is set to off at the factory. Utility Display Screen Saver Mode...
Getting Started Configuring for Remote Control Configuring for Remote Control GPIB Interface Configuration Utility GPIB/RS-232 LAN GPIB Address 1. Press > > Enter 2. Use the numeric keypad to set the desired address and press The arrow keys or the front panel knob can be used to set the desired address. The signal generator’s GPIB address is set to 19 at the factory.
Page 15
Getting Started Configuring for Remote Control IP Address 6. Press and enter a desired address. Use the left and right arrow keys to move the cursor. Use the up and down arrow keys, front panel knob, Clear Text or numeric keypad to enter an IP address. To erase the current IP address, press the softkey.
Getting Started Configuring for Remote Control RS-232 Interface Configuration Utility > GPIB/RS-232 LAN > RS-232 Setup 1. Press RS-232 Baud Rate 2. Press 3. Press the desired baud rate softkey. RS-232 Echo Off On 4. Press This toggles the state of the SCPI echoing on the RS-232 connection. Set as desired. Reset RS-232 5.
Getting Started Ordering Accessories Ordering Accessories The following accessories can be ordered when a signal generator is purchased, or at any time afterward. To order accessories, refer to “Contacting Agilent Sales and Service Offices” on page Front Handles and Rack Mount Flanges Handles can be purchased and attached to the front of the signal generator.
Page 18
Getting Started Ordering Accessories Table 2-3 Available E8257D/67D PSG Documentation Document Type Description Part Number E8257D Data Sheet • 5989-0698EN available options • warranted specifications and typical performance E8267D Data Sheet • 5989-0697EN available options • warranted specifications and typical performance User’s Guide •...
Getting Started Proper Usage and Cleaning Proper Usage and Cleaning The signal generator cover protects against physical contact with internal assemblies that contain hazardous voltages, but does not protect against the entrance of water. To avoid damage and personal injury, ensure that liquid substances are positioned away from your signal generator.
Getting Started Contacting Agilent Sales and Service Offices Contacting Agilent Sales and Service Offices Assistance with test and measurements needs and information on finding a local Agilent office are available on the Internet at: http://www.agilent.com/find/assist You can also purchase E8257D/67D PSG accessories or documentation items on the Internet at: http://www.agilent.com/find/psg If you do not have access to the Internet, please contact your field engineer.
Returning a Signal Generator to Agilent Technologies Returning a Signal Generator to Agilent Technologies To return your signal generator to Agilent Technologies for servicing, follow these steps: 1. Gather as much information as possible regarding the signal generator’s problem. 2. Call the phone number listed on the Internet (http://www.agilent.com/find/assist) that is specific to your geographic location.
Page 22
Getting Started Returning a Signal Generator to Agilent Technologies Chapter 2 Download from Www.Somanuals.com. All Manuals Search And Download.
Operation Verification Operation verification is a series of tests that, when completed, will either ensure that the signal generator is operating correctly, or will assist in pointing to the problem area. Operation verification does not ensure performance to specifications, but should provide a level of confidence that the signal generator is operating correctly within a minimum amount of time.
Operation Verification Performing a Self-Test Performing a Self-Test The self-test is a series of internal tests that checks different signal generator functions. If this test fails, refer “Self-Test Failure” on page 21 for further instructions. Perform the following procedure to run a self-test: 1.
2. If the self-test continues to fail, send the signal generator to an Agilent service center for repair, with a detailed description of the failed test(s) and any other error messages that appeared on the display. “Returning a Signal Generator to Agilent Technologies” on page 17 for return instructions.
Operation Verification Checking the Maximum Leveled Power Checking the Maximum Leveled Power Perform the following procedure to check the maximum leveled power: Preset 1. Press 2. Attach a 50Ω load to the RF OUTPUT. A power sensor, attenuator, or 50Ω termination is an example of a 50Ω...
Page 27
Operation Verification Checking the Maximum Leveled Power 3. Ensure that the RF OUTPUT connector is connected to a 50Ω load. 4. Ensure that the power level entered corresponds to the value listed in Table 3-1. 5. Refer to Table 3-2 for the recommended equipment and measure the output power.
Page 28
Operation Verification Checking the Maximum Leveled Power Table 3-1 Frequency and Power Level Limits Instrument Option Amplitude Start Stop # Points Standard Option 1EA Frequency Frequency E8257D 11 dBm 15 dBm 250 kHz 3.2 GHz Option 520 11 dBm 18 dBm 3.21 kHz 20 GHz With Option 1E1...
Operation Verification Checking the Output Power Checking the Output Power This test verifies the CW output power from the signal generator. If this test fails, refer to “Problems with Output Power” on page 32 for further instructions. Table 3-2 shows the recommended equipment for use with the E8257D signal generator when performing this test.
Page 30
Operation Verification Checking the Output Power Table 3-2 Recommended Equipment for the E8257D E8257D E8257D E8257D E8257D Recommended Agilent Option Option Option Option Test Equipment Test Equipment Adapter, Agilent 1250-1749 APC 3.5 (f) to 3.5 (f) to (f) (f) adapter Adapter, Agilent 1250-1744 3.5 (f) to...
Page 31
Operation Verification Checking the Output Power Table 3-3 Recommended Equipment for the E8267D E8267D E8267D E8667D Recommended Agilent Option Option Option Test Equipment Test Equipment Adapter, Agilent 1250-1749 APC 3.5 (f) to APC 3.5 (f) to (f) adapter Adapter, Agilent 1250-1744 APC 3.5 (f) to APC 3.5 (f) to Type-N (m) adapter Type-N (m)
Page 32
Operation Verification Checking the Output Power Perform the following procedure to check the signal generator output power: 1. Zero and calibrate the power sensor to the power meter as shown: 2. Connect the equipment as shown: Preset 3. Press RF On/Off 4.
Page 33
Operation Verification Checking the Output Power Table 3-4 Leveled Output Power Limits, E8257D E8257D Option 520 Option 540 Frequency Amplitude Limits Frequency Amplitude Limits Option Option 250 kHz None ±0.6 250 kHz None ±0.6 ±0.6 ±0.6 ±0.6 ±0.6 1E1 & 1EA ±0.6 1E1 &...
Page 34
Operation Verification Checking the Output Power Table 3-5 Leveled Output Power Limits, E8257D cont. E8257D Option 550 Option 567 Frequency Amplitude Limits Frequency Amplitude Limits Option Option 250 kHz None ±0.6 250 kHz None ±0.6 ±0.6 ±0.6 ±0.6 ±0.6 1E1 & 1EA ±0.8 1E1 &...
Page 35
Operation Verification Checking the Output Power Table 3-5 Leveled Output Power Limits, E8257D cont. Option 550 Option 567 Frequency Amplitude Limits Frequency Amplitude Limits Option Option 60 GHz None ±1.2 60 GHz None ±1.2 ±1.2 ±1.2 ±1.2 ±1.2 1E1 & 1EA ±1.2 1E1 &...
Operation Verification Checking the Output Power Problems with Output Power If a problem occurs when verifying output power levels, perform the following procedure: 1. Verify the power sensor being used in this test is the appropriate sensor. 2. Verify the power sensor calibration factors, if required, are correct. 3.
16.) Certification Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by the Institute’s calibration facility, and to the calibration facilities of other International Standards Organization members.
Regulatory Information Compliance with Canadian EMC Requirements Compliance with Canadian EMC Requirements This is to declare that this ISM device complies with Canadian ICES-001. (Cet appareil ISM est conforme a la norme NMB du Canada.) Chapter 4 Download from Www.Somanuals.com. All Manuals Search And Download.
Page 40
Regulatory Information Compliance with Canadian EMC Requirements Chapter 4 Download from Www.Somanuals.com. All Manuals Search And Download.
Page 41
Index AC power cord, 6, IEC Publication 61010, AC symbol, instruction manual symbol, accessories, inverse video adjustment, address, GPIB IP address, setting, altitude requirements, ISM1-A symbol, assistance, Australian Communications Authority (C-tick) mark, kits, brightness adjustment, LAN configuration, light adjustment, Canadian EMC requirements, Canadian Standards Association (CSA) mark, CE mark, maintenance agreements,...
Page 42
Index troubleshooting maximum leveled power, output power, self-test, ventilation requirements, verification, operation, Index Download from Www.Somanuals.com. All Manuals Search And Download.
Page 43
Free Manuals Download Website h p://myh66.com h p://usermanuals.us h p://www.somanuals.com h p://www.4manuals.cc h p://www.manual-lib.com h p://www.404manual.com h p://www.luxmanual.com h p://aubethermostatmanual.com Golf course search by state h p://golfingnear.com Email search by domain h p://emailbydomain.com Auto manuals search h p://auto.somanuals.com TV manuals search h p://tv.somanuals.com...