Custom Arb Waveform Generator
Working with Modulation Types
2. Press
>
Mode
Custom
>
Rows
Confirm Delete All Rows
This loads a default 4QAM I/Q modulation and clears the I/Q Values editor.
3. Enter the I and Q values listed in the following table:
Symbol
Data
0000
0
0001
1
0010
2
0011
3
a. Press
>
0.5
Enter
b. Press
>
.
1
Enter
c. Enter the remaining I and Q values.
As the I value updates, the highlight moves to the first Q entry (and provides a default value of
0) and an empty row of data appears below the first row. As the Q value updates, the highlight
moves to the next I value. As you press the numeric keys, the numbers display in the active entry
area. If you make a mistake, use the backspace key and retype.
Also note that 0.000000 appears as the first entry in the list of Distinct Values, and that
0.500000 and 1.000000 are listed as the distinct values.
>
4. Press
More (2 of 2)
An I/Q State Map is displayed from the current values in the I/Q Values table.
The I/Q State Map in this example has four symbols. The I/Q State Map uses the following four
unique values: 0.5, 1.0, −0.5, and −1.0 to create the four symbols. It is not the number of values
that defines how many symbols a map has, but how those values are combined.
5. Press
.
Return
When the contents of an I/Q Values table have not been stored, I/Q Values (UNSTORED)
appears on the display.
6. Press
>
More (1 of 2)
If there is already a file name from the Catalog of IQ Files occupying the active entry
area, press the following keys:
7. Enter a file name (for example, NEW4QAM) using the alpha keys and the numeric keypad.
8. Press
. The user–defined I/Q State Map should now be stored in the Catalog of IQ Files
Enter
and can be recalled even after the signal generator has been turned off.
174
>
Real Time I/Q Baseband
.
I Value
0.500000
−0.500000
0.500000
−0.500000
.
.
Display I/Q Map
>
.
Load/Store
Store To File
Editing Keys
>
>
Modulation Type
Define User I/Q
Q Value
1.000000
1.000000
−1.000000
−1.000000
>
Clear Text
>
>
More (1 of 2)
Delete All
Chapter 6