Optimizing Performance
Creating and Applying User Flatness Correction
Figure 4-4
Configure the Signal Generator
1. Press
.
Preset
2. Configure the signal generator to interface with the power meter.
a. Press
Amplitude
.
E4419B
b. Press
Meter Address
c. For E4417A and E4419B models, press
channel.
d. Press
Meter Timeout
if unsuccessfully attempting to communicate with the power meter.
3. Press
>
More (2 of 2)
This opens the User Flatness table editor and presets the cal array frequency/correction list.
4. Press
Configure Step Array
This opens a menu for entering the user flatness step array data.
5. Press
>
Freq Start
1
6. Press
>
Freq Stop
10
136
User Flatness Correction Equipment Setup
>
>
More (1 of 2)
User Flatness
> enter the power meter's GPIB address >
to adjust the length of time before the instrument generates a timeout error
>
Configure Cal Array
More (1 of 2)
.
>
.
GHz
>
.
GHz
>
>
More (1 of 2)
Power Meter
to select the power meter's active
Meter Channel A B
>
>
Preset List
Confirm Preset
>
,
,
E4416A
E4417A
E4418B
.
Enter
.
, or