Agilent Technologies E8257D PSG User Manual page 158

Psg signal generators
Hide thumbs Also See for E8257D PSG:
Table of Contents

Advertisement

Optimizing Performance
Creating and Applying User Flatness Correction
NOTE
For specific frequency/amplitude ranges, see the mm–wave source module specifications.
2. Configure the signal generator to interface with the power meter.
a. Press
Amplitude
.
E4419B
b. Press
Meter Address
c. For E4417A and E4419B models, press
channel.
d. Press
Meter Timeout
if unsuccessfully attempting to communicate with the power meter.
3. Press
>
More (2 of 2)
This opens the User Flatness table editor and resets the cal array frequency/correction list.
4. Press
Configure Step Array
This opens a menu for entering the user flatness step array data.
5. Press
>
Freq Start
26.5
6. Press
>
Freq Stop
40
7. Press
>
# of Points
28
This enters the desired flatness–corrected frequencies (26.5 GHz to 40 GHz in 500 MHz intervals)
into the step array.
8. Press
>
Return
Load Cal Array From Step Array
This populates the user flatness correction array with the frequency settings defined in the step
array.
9. Press
>
Amplitude
0
10. Press
.
RF On/Off
This activates the RF output and the RF ON annunciator is displayed on the signal generator.
Perform the User Flatness Correction
NOTE
If you are not using an Agilent E4416A/17A/18B/19B power meter, or if your power meter
does not have a GPIB interface, you can perform the user flatness correction manually. For
instructions, see
1. Press
>
More (1 of 2)
This creates the user flatness amplitude correction value table entries. The signal generator begins
the user flatness correction routine and a progress bar is shown on the display.
2. When prompted, press
This loads the amplitude correction values into the user flatness correction array.
If desired, press
Configure Cal Array
This opens the user flatness correction array, where you can view the list of defined frequencies
144
>
More (1 of 2)
>
User Flatness
> enter the power meter's GPIB address >
to adjust the length of time before the instrument generates a timeout error
>
Configure Cal Array
More (1 of 2)
.
>
.
GHz
>
.
GHz
>
.
Enter
>
>
.
dBm
Performing the User Flatness Correction Manually
>
.
User Flatness
Do Cal
.
Done
.
>
More (1 of 2)
>
Power Meter
to select the power meter's active
Meter Channel A B
>
>
Preset List
Confirm Preset
Confirm Load From Step Data
>
E4416A
,
E4417A
.
Enter
.
.
below.
,
E4418B
, or
Chapter 4

Hide quick links:

Advertisement

Table of Contents
loading