6.3.7 E71: PB T°
The temperature of the radiator of the PowerBox has become too high. Make sure there is enough room
for air circulation, move the PowerBox to a cooler or shaded area and allow it to cool down in stand-by
mode before attempting to make further exposures.
6.3.8 E72: PB T°
The temperature inside the PowerBox has become too high. Make sure there is enough room for air
circulation, move the PowerBox to a cooler or shaded area and allow it to cool down in stand-by mode
before attempting to make further exposures.
6.3.9 E80
THROUGH
One of the lamps controlled by the PowerBox is defective. Replace the lamp or disable the lamp control if
you are authorized to do so.
6.3.10
E87, E88:
The water or oil flow control circuit detects an error. Please check the cooling circuit or disable the control
if you are not using liquid cooling.
6.3.11
E89:
The external buzzer is defective. Replace it or disable the buzzer control if you are authorized to do so.
6.3.12
E90
THROUGH
One of the 3 door controls is giving an error. Check the door contact or disable the control if you are
authorized to do so.
6.3.13
E95:
The red lamp located on the generator is defective. Replace the lamp or disable the lamp control if you are
authorized to do so.
6.3.14
E97:
The key (12) is not present or is in the X0 position. Set it to the XI position in order to enable X-rays.
6.3.15
E115:
The PowerBox cannot communicate with the generator. Make sure the interconnection cable is connected
properly or try to replace it.
RAD MAX
AMB MAX
E86:
DEFECTIVE LAMP
WATER OR OIL FLOW DEFECT
DEFECTIVE EXTERNAL BUZZER
E92:
DOOR OPEN
DEFECTIVE GENERATOR LAMP
KEY OFF
NO GENERATOR