Guard Rings; Carousel/Diaphragm (Not For Cp300C) - Teledyne CP Series User Manual

Table of Contents

Advertisement

User Manual
CP Series
Version 3.1 – June 2020
3.1.2 G
UARD RINGS
The guard rings provide handles to easily manipulate and position the generator but more importantly they
provide shock absorption in case the generator is dropped.
3.1.3 C
AROUSEL
There are 5 positions available on the carousel.
4 Positions are standard: Shutter, Laser pointer, Beryllium window, and
full open with 3mm aluminum filter. The 5
size desired but is populated with a 480x100 window by standard if no
other size was specified during the order process.
You can switch between positions by simply rotating the carousel
wheel. The generator will automatically detect which carousel position
is selected and communicate this information to the PowerBox.
No exposure will be possible if the carousel is set between two
positions, or if it is set in the pointer position.
No exposure is possible above 50kV if the Beryllium window is set.
A warning will be displayed on the PowerBox if the shutter is not set
during the preheating, or if it is set during a normal exposure.
The laser pointer can be switched on and off by pressing the red power button for more than one second.
It will switch off automatically after 20 seconds. The laser pointer is powered by a 3 Volt coin battery which
can be replaced at the back of the laser pointer module. Please refer to the Maintenance chapter on further
information on how to change the coin battery.
Carousel position
/D
(
IAPHRAGM
NOT FOR
position can fit any window
th
Function
Shutter
The shutter is a lead cover which closes the X-ray output. Use this
position to reduce the radiation produced during preheating.
Full open
A 3mm aluminum filter is applied for the full 60°x40° opening of the
X-ray output.
CP300C)
17

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents