Teledyne CP Series User Manual page 66

Table of Contents

Advertisement

User Manual
CP Series
Version 3.1 – June 2020
Command description
Command syntax
Description
?SE
Ask for actual shot values
#kV<XXX>
Program new kV value
#MA<XXX>
Program new mA value
Fields description
Reply from PowerBox
Field name
<XXX>
!SE<XXX>,<YYY>,<ZZZ>
<YYY>
<ZZZ>
<XXX>
<YYY>
<E>
!kV<YYY>,<E>,<AAAA>
<AAAA>
<XXX>
<YYY>
<E>
!MA<YYY>,<E>,<AAA>
<AAA>
Description
Programmed kV setup value
Programmed mA setup value
Programmed time setup value
Requested new kV setup value
Programmed kV setup value
Error code:
0: no error (<YYY>=<XXX>)
1: no generator connected
2: generator rental date is over
3: requested kV value is below min kV
4 requested kV value is above max kV
5: kV ok but mA has to be adjusted
6 : preheat is required to shoot at this kV value
Argument depending of <E>:
If <E> = 0,1,2  0
If <E> = 3 min kV value for this generator
If <E> = 4 max kV value for this generator
If <E> = 5 adjusted mA value
If <E> = 6 required preheat time
Requested new mA setup value
Programmed mA setup value
0: no error
1: no generator connected
2: generator rental date is over
3: requested mA value is below min mA
4: requested mA value is above max mA
5: preheat required – mA set automatically
If <E>=0,1,2,5  0
If <E> = 3  min mA value
If <E> = 4  max mA value
units
kV
mA/10
s
kV
kV
N.A.
N.A.
kV
kV
mA
s
mA
mA
N.A.
N.A.
mA
mA
66