Removal/ Installation, Replacement and Adjustment
RF Oscillator
1 Remove Torch by pulling down and out.
2 Remove 4 torx screws from RF Oscillator cover and remove cover.
3 Loosen blue fibre optic sensor screw by rotating thimble anticlockwise and remove fibre
optic cable from receptacle.
4 Place absorbent tissue into base of RF compartment to catch water overspill that will occur
when hoses are disconnected.
5 Ensure water supply is turned off. Disconnect lower water hose first by pressing stainless
steel latch and pulling connector to the left.