Agilent Technologies 5110 ICP-OES Service Manual page 269

Hide thumbs Also See for 5110 ICP-OES:
Table of Contents

Advertisement

10
Installation Guide (5110 ICP-OES)
Hardware installation
The Agilent 5110 ICP-OES instrument weighs 180Kg crated and 106Kg uncrated and should
WARNING
not be moved without the correct lifting equipment. Not following correct procedure for lifting
heavy equipment could result in injury.
1 Remove straps and clips from crate and remove walls of crate to expose instrument on
pallet.
2 Use a 12mm wrench to loosen the 4 shipping bolts and remove pallet clamps by pressing
down and sliding away from instrument.
3 Using suitably rated lifting equipment place instrument onto bench.
4 With instrument resting on the bench use handling points to slide instrument into position.
The FSE is required to connect the available services and peripherals.
Agilent 5110/5100 ICP-OES Service Manual
Agilent Restricted
269