4
Removal/ Installation, Replacement and Adjustment
Torch loader micro switch
1 Remove 2 hex head screws with 5mm hex driver.
2 Slide plate away from instrument to expose micro switch interlock.
3 Remove torx 10 screw.
4 Disconnect loom and replace switch as required. Be sure to tuck loom out of the way during
reassembly.