Agilent Technologies 5110 ICP-OES Service Manual page 131

Hide thumbs Also See for 5110 ICP-OES:
Table of Contents

Advertisement

4
Removal/ Installation, Replacement and Adjustment
Polychromator
1 Remove the pilot mirror casting access cover.
2 Remove the slit access plug using 3mm hex driver to remove 2 hex screws.
3 Remove lower slit retaining screw.
4 Remove upper slit retaining screw.
5 Remove the slit assembly away from the casting and through the slit access in the top of the
casting being careful not to touch the pilot mirror on the way out.
Agilent 5110/5100 ICP-OES Service Manual
Agilent Restricted
131