Agilent Technologies 5110 ICP-OES Service Manual page 201

Hide thumbs Also See for 5110 ICP-OES:
Table of Contents

Advertisement

7
Service Diagnostics and Fault Finding
ICP Expert Diagnostics
Note graph representing RFPS input voltage (orange) and RFPS current before, during and after
ignition then varying power settings.
After plasma stabilization FET bias voltages 1 and 2 settle to ~1.9V. RF DC power supply settles
to ~194V @8.2A for a power setting of 1.2kW.
Frequency settles to ~26.5MHz and will change depending on whether a solution is being
aspirated and the makeup of the solution being aspirated.
Plasma run status typical Gas flows and pressures (Poly boost purge and cone purge on also).
Agilent 5110/5100 ICP-OES Service Manual
Agilent Restricted
201