7
Service Diagnostics and Fault Finding
Configure Instrument IP Address
2 Click on "Edit Network Configuration" to change the IP information and when finished click on
"Apply".
If the instrument IP address is changed the connected PC IP address will need to also be
NOTE
changed to at least match the first 3 groups of numbers, i.e.; 192.168.1.xxx for communication
to be maintained. Preferably enter the 4
(192.168.1.127).
3 Enter the word "admin" and the word "Password" to save changes.
The instrument status will show "Idle" if the instrument connection is correct.
192
Agilent Restricted
th
group to be within a couple of the Instrument, i.e.;
Agilent 5110/5100 ICP-OES Service Manual