10
Installation Guide (5110 ICP-OES)
Analytical Tests
Precision
The precision for the 10 replicates of sample readings on the multi element standard should be
less than the following:
Table 5
Element
Al167, As, Pb & Se
All other lines
Agilent 5110/5100 ICP-OES Service Manual
Agilent 5110 ICP-OES Instruments
Specification
2.6 %RSD
1.5 %RSD
Agilent Restricted
289