Agilent Technologies 5110 ICP-OES Service Manual page 259

Hide thumbs Also See for 5110 ICP-OES:
Table of Contents

Advertisement

9
iTroubleshoot/Parts Predict
Problem Solving Table
Table 1
Symptom Area
Problem
Gas Control
Gas Control Ready LED Red
Gas Control
Gas Control Ready LED Red
Gas Control
Gas Control LED Red
Instrument Control
Instrument won't initialize
Instrument Control
Instrument won't initialize
Instrument Control
Instrument won't initialize
Instrument Control
Instrument won't initialize
Instrument Control
Electronics Ready LED Red
Instrument Power
Soft Switch LED flashing at 4Hz
Instrument Power
Soft Switch LED flashing at 4Hz
Instrument Power
Soft Switch LED off
Instrument Power
Soft Switch LED off
Instrument Status LED
Status LED Flashing Yellow
Instrument Status LED
Status LED Red
Instrument Status LED
Status LED Yellow
Instrument Status LED
Status LED Yellow
Low Intensities
Low Intensities at high/Low
Wavelengths (May not
Calibrate)
Low Intensities
Low Intensities at high/Low
Wavelengths (May not
Calibrate)
Low Intensities
Low Intensities at high/Low
Wavelengths (May not
Calibrate)
Low Intensities
Low Intensities at high/Low
Wavelengths (May not
Calibrate)
Low Intensities
Low Intensities at high/Low
Wavelengths (May not
Calibrate)
Low Intensities
Low Intensities at high/Low
Wavelengths (May not
Calibrate)
Low Intensities
Low Intensities at high/Low
Wavelengths (May not
Calibrate)
Low Intensities
Low Intensities at high/Low
Wavelengths (May not
Calibrate)
Low Intensities
Low Intensities at high/Low
Wavelengths (May not
Calibrate)
Agilent 5110/5100 ICP-OES Service Manual
Problem Description
Gas Control not calibrated
Plasma Gas Flows incorrect Replace Plasma Gas Module
Gas Control Module Fault
Firmware Error
Firmware Error
Control Board Fault
MPM Failed
Communications Failure
MPM Failed
Power Supply inadequate
Instrument Circuit Breaker
Mains Power Supply issue
Host not connected
Firmware Error
Interlock not satisfied
Interlock not satisfied
Argon Ratio incorrect
Argon Ratio incorrect
Argon Ratio incorrect
High Dark Current values
High Dark Current values
Leaking fitting
Parameters not optimized
Sample Delivery issue
Sample Intro blockage
Solution
Replace Gas Control Module
Replace Gas Control Module
Replace P500 PCA
Reload Firmware
Replace Instrument Control PCA G8010-65200
Replace MPM
Replace Instrument Control PCA G8010-65200
Replace
Verify Power Supply
Check Circuit Breaker On
Verify Customer Power Supply
Check for Communication to PC
Reload Firmware
Check RF Door, Torch Handle,
Water Flow
Replace Instrument Control PCA G8010-65200
Allow instrument to warm up for
4 hours
Polychromator temperature
control failing - (Optics Ready
LED red section)
Pre Optics or Polychromator
requires realignment
Recalibrate Detector - Check for
light leak, shutter sticking
Realign Optics, Check Camera
offsets using PolyAlign S/W
Check Nebulizer, Pump tubing,
Sampler, Bubbles in Tube
Optimize Nebulizer flow,
Viewing height, Plasma Power,
Pump Speed
Check Pump tube tension or
stretched pump tube
Nebulizer Blocked - Clean
Agilent Restricted
Part Number(s)
G8010-60800
G8010-60826
G8010-60800
G8010-65204
G8010-67300
G8010-67300
3710034600,
3710046900
G8010-60255
259

Advertisement

Table of Contents