Agilent Technologies 5110 ICP-OES Service Manual page 63

Hide thumbs Also See for 5110 ICP-OES:
Table of Contents

Advertisement

4
Removal/ Installation, Replacement and Adjustment
Tools and Supplies for 5110 instrument
Tool
Adjustable wrench (200mm)
Flow Meter
Digital Multimeter
ESD Strap
Plastic tubing cutter
Blades for plastic tube cutter
Mirror
5ppm WaveCal solution
Blank Solution
Be aware of the following precautions:
CAUTION
• Carry out disassembly/assembly work in a clean dust- free location.
• Secure sufficient space and confirm that the surroundings are safe before starting.
• Select and use tool(s) suitable for the job at hand.
• Use tools that are clean and degreased.
• Where it is necessary to make alignment marks, do so on sections that do not affect the
functioning of the part.
• When removing wiring etc., make a note of the condition, routing, connections, etc., before
proceeding.
• Do not excessively twist or strike the parts with tools such as a driver, etc.
• Handle removed parts with sufficient care; keep them in a clean location.
• Organize removed parts.
• Clean or wash the parts as necessary.
• When completing work, confirm that the wiring is connected in its original layout.
• Always make sure you are properly grounded (e.g., wearing an ESD strap) before handling
CAUTION
electro-static sensitive components such as printed circuit boards.
• As electric circuits may short, remove any metal jewelry such as rings, etc. before
proceeding.
• When conducing work related to the gas piping, water hose, inspect for damage and leaks.
• If a part (such as O-rings, etc.) is found to be deteriorated or damaged when removing,
replace it.
• When unplugging and plugging connectors, make sure not to twist them. Twisting them
may lead to contact failure.
• When unplugging connectors, support the wiring and do not pull on the wiring.
Agilent 5110/5100 ICP-OES Service Manual
Part Number
Where used
Water fittings, General
Gas flows
Fault Finding
7910031300
PCA handling
8710-1930
Hoses
8710-1931
707-0027
Inspection of hard to reach places
6610030100
For calibration of wavelengths
5190-7001
For performance testing
Agilent Restricted
63