Agilent Technologies 5110 ICP-OES Service Manual page 263

Hide thumbs Also See for 5110 ICP-OES:
Table of Contents

Advertisement

9
iTroubleshoot/Parts Predict
Problem Solving Table
Table 1
Symptom Area
Problem
Poor Precision
Poor Precision across multiple
wavelengths
Poor Precision
Poor Precision across multiple
wavelengths
Pre Optics
Pre optics Ready LED Red
Pre Optics
Pre optics Ready LED Red
Pre Optics
Pre optics Ready LED Red
Pre Optics
Pre optics Ready LED Red
Pre Optics
Pre optics Ready LED Red
Pre Optics
Pre optics Ready LED Red
RF Control
RF Ready LED Red
Signal Drift
Signal drift - Thermal Stabilizer
issue
Signal Drift
Signal drift - Polychromator
temperature unstable
Signal Drift
Signal drift - Thermal Stabilizer
issue
Signal Drift
Signal drift - Thermal Stabilizer
issue
Signal Drift
Signal Drift -Exhaust Extraction
inadequate
Torch Loader
Torch Loader Ready LED Red
Torch Loader
Torch Loader Ready LED Red
Torch Loader
Torch Loader Ready LED Red
Water Cooling
Water Cooling Ready LED Red
Water Cooling
Water Cooling Ready LED Red
Water Cooling
Water Cooling Ready LED Red
Water Cooling
Water Cooling Ready LED Red
Water Cooling
Water Cooling Ready LED Red
Water Cooling
Water Cooling Ready LED Red
Water Cooling
Water Cooling Ready LED Red
Water Cooling
Water Cooling Ready LED Red
Water Cooling
Water Cooling Ready LED Red
Agilent 5110/5100 ICP-OES Service Manual
Problem Description
Wavelengths off peak
Wavelengths off peak
Beam Combiner Fault
Beam Combiner Fault
Radial Scan Fault
Radial Scan Fault
Beam Combiner Fault
Beam Combiner Fault
RF Comms Error
Not at 35°C
Poly not at 35°C
Not driving to 35°C
Not driving to 35°C
Exhaust flow too low
Torch interlock fault
Torch interlock fault
Torch not loaded correctly
CCD Water Flow out of
Spec
CCD Water Flow out of
Spec
Low Flows for CCD and RF
SSRF Solenoid Failed
SSRF Solenoid Failed
SSRF Water Flow out of
Spec
SSRF Water Flow out of
Spec
Water supply Pressure
Water Temp out of spec
Solution
Instrument warm up for 4 hours
Perform Instrument Calibration
Replace Beam combiner Opto
sensor
Replace Pre Optics Motor
Control PCA
Replace radial Scan opto sensor G8010-65208
Replace Pre Optics Motor
Control PCA
Replace Beam combiner Opto
sensor
Replace Pre Optics Motor
Control PCA
Replace SSRF Control PCA
Wait for stabilization
Wait for stabilization
(No current showing in
diagnostics) Replace Thermal
stabilizer
Replace Mains Power Module
(MPM)
Check for correct flow using
tissue
Adjust or replace interlock
switch
Replace Instrument Control PCA G8010-65200
Load Torch
Replace Water Flow Sensor
Replace Instrument Control PCA G8010-65200
Blocked Water Filter - Clean or
replace
Replace Solenoid
Replace Instrument Control PCA G8010-65200
Clear restriction
Replace SSRF Flow Sensor
Check and Adjust
Check Chiller and adjust
Agilent Restricted
Part Number(s)
G8010-65208
G8010-65206
G8010-65206
G8010-65208
G8010-65206
G8010-65205
G8010-60605
G8010-67300
G8010-60226
110796700
G1833-66024
G8400-60342
110796700
263

Advertisement

Table of Contents
loading

This manual is also suitable for:

5100 icp-oes

Table of Contents