Agilent Technologies 5110 ICP-OES Service Manual page 260

Hide thumbs Also See for 5110 ICP-OES:
Table of Contents

Advertisement

9
iTroubleshoot/Parts Predict
Problem Solving Table
Table 1
Symptom Area
Problem
Low Intensities
Low Intensities at high/Low
Wavelengths (May not
Calibrate)
Low Intensities
Low Intensities at high/Low
Wavelengths (May not
Calibrate)
Low Intensities
Low Intensities at high/Low
Wavelengths (May not
Calibrate)
Low Intensities
Low Intensities at high/Low
Wavelengths (May not
Calibrate)
Low Intensities
Low Intensities at Low
Wavelengths (May not
Calibrate)
Low Intensities
Low Intensities at Low
Wavelengths (May not
Calibrate)
Low Intensities
Low Intensities at Low
Wavelengths (May not
Calibrate)
Low Intensities
No Signal
Low Intensities
No Signal
Low Intensities
No Signal
Low Intensities
No Signal
Low Intensities
Random Wavelengths
intensities failing
Low Intensities
Random Wavelengths
intensities failing
Low Intensities
Random Wavelengths
intensities failing
Low Intensities
Random Wavelengths
intensities failing
Low Intensities
Random Wavelengths
intensities failing
Low Intensities
Random Wavelengths
intensities failing
Low Intensities
Random Wavelengths
intensities failing
Low Intensities
Air Present in optics
260
Agilent Restricted
Problem Description
Sample Intro blockage
Sample Intro blockage
Wavelengths Off Peak
Wrong Std Concentration
Blocked Purge Filter
Purge time insufficient
Windows Contaminated
Spraychamber not misting
Spraychamber not misting
Spraychamber not misting
Spraychamber not misting
Contaminated Blank
High Dark Current values
High Dark Current values
Moisture on Detector Lens
Sample Matrix issue
Wavelengths Off Peak
Wavelengths Off Peak
Ensure Purging is on
Agilent 5110/5100 ICP-OES Service Manual
Solution
Tubing kinked, blocked, dirty -
Rectify
Torch injector blocked - Clean
Recalibrate Instrument
Prepare new standards
Replace Purge Filter
Purge for 4 hrs minimum and
ensure all purges are on
Replace Axial/Radial Windows
No Nebulizer gas due to fitting
not pushed home - Reseat
Nebulizer Blocked - Clean or
replace
No uptake due to wrong pump
pressure - adjust
Pump tubes stretched - Replace
Prepare new blank
Recalibrate Detector - Check for
light leak, shutter sticking
Realign Optics, Check Camera
offsets using PolyAlign S/W
Moisture in purge gas supply
(Argon) - Switch off chiller for
30mins to verify. Check with
Polyalign/LED on, no plasma
and 10ms integrate. There
should be no peaks, especially,
in the region 330-180nm (DLA
43 to 35)
Interference from EIE's. Change
to Radial view
Recalibrate Instrument
Allow 4 hours for instrument
warmup
Optics Purge, Boost Purge, Cone
Purge, Snout Purge all on
Part Number(s)
2410020500
G8010-60228
6610030100
G8010-60136
G8010-68014,
G8010-68015
G8010-60255
5190-7001