Agilent Technologies 5110 ICP-OES Service Manual page 262

Hide thumbs Also See for 5110 ICP-OES:
Table of Contents

Advertisement

9
iTroubleshoot/Parts Predict
Problem Solving Table
Table 1
Symptom Area
Problem
Plasma Not igniting
Won't ignite at high altitude
above 1500m
Plasma Not Sustained
Plasma Looker not seeing
plasma
Poor Precision
Poor Precision across a select
few wavelengths
Poor Precision
Poor Precision across a select
few wavelengths
Poor Precision
Poor Precision across multiple
wavelengths
Poor Precision
Poor Precision across multiple
wavelengths
Poor Precision
Poor Precision across multiple
wavelengths
Poor Precision
Poor Precision across multiple
wavelengths
Poor Precision
Poor Precision across multiple
wavelengths
Poor Precision
Poor Precision across multiple
wavelengths
Poor Precision
Poor Precision across multiple
wavelengths
Poor Precision
Poor Precision across multiple
wavelengths
Poor Precision
Poor Precision across multiple
wavelengths
Poor Precision
Poor Precision across multiple
wavelengths
Poor Precision
Poor Precision across multiple
wavelengths
Poor Precision
Poor Precision across multiple
wavelengths
Poor Precision
Poor Precision across multiple
wavelengths
Poor Precision
Poor Precision across multiple
wavelengths
Poor Precision
Poor Precision across multiple
wavelengths
Poor Precision
Poor Precision across multiple
wavelengths
Poor Precision
Poor Precision across multiple
wavelengths
Poor Precision
Poor Precision across multiple
wavelengths
262
Agilent Restricted
Problem Description
Optimize ignition gas flows
SSRF Control PCA faulty
Wavelengths Off Peak
Wavelengths Off Peak
Electronic Noise
High Salts
Inconsistent sample
delivery
Inconsistent sample
delivery
Inconsistent sample
delivery
Inconsistent sample
delivery
Low 1st Replicate (flyer)
Optimization of parameters
Optimization of parameters
Previously running organics
Running partially soluble
(salts)
Sample intro contamination
Sample intro contamination
Sample intro contamination
Sample intro contamination
Unstable plasma
Unstable plasma
Unstable plasma
Agilent 5110/5100 ICP-OES Service Manual
Solution
Using Trial and error set correct
ignition gas flows (plasma and
auxiliary) for thinner air at higher
altitude
Replace SSRF Control PCA
Instrument warm up for 4 hours
Recalibrate Instrument
Camera, / Control PCA, Fault
(Investigate)
Partial Blocking of Nebulizer and
injector tube. - Clean
Pump Tube Pressure incorrect -
Adjust
Pulsations due to low pump
speed - Change
Crimped or damaged pump
tube - Change
Uptake delay time too short for
sample uptake tubing length
(first reading low). - Increase
Increase Delay time to align with
sample tube length
Optimize Nebulizer Flow
Optimize Viewing height (Radial)
Contaminated Spraychamber,
Nebulizer, Torch injector - Clean
Partial blocking of nebulizer and
injector tube - Clean
Spraychamber walls dirty
(beading) - Clean
Nebulizer partially blocked
(uneven spray) - Clean
Nebulizer tip damaged - Replace
Torch injector tube damaged or
partially blocked - Clean
Contaminated Argon - Check
and replace
Air Leak - Investigate Optics
seals (Rubber grommets)
RF system failing ("Plasma not
igniting" section)
Part Number(s)
G8010-65205
G8014-60100
G8010-60255
3710034600,
3710046900
G8010-60255
G8010-60256
G8010-60255
G8010-60255
G8010-60228