Agilent Technologies 5110 ICP-OES Service Manual page 135

Hide thumbs Also See for 5110 ICP-OES:
Table of Contents

Advertisement

4
Removal/ Installation, Replacement and Adjustment
Polychromator
15 Remove the screws retaining the camera connection board to the casting.
16 Remove the 2 screws that secure the D range connector to the board.
17 Using a 7 mm nut driver, remove the three Nylock nuts that hold the camera assembly in the
casting.
18 Carefully remove the camera assembly.
Agilent 5110/5100 ICP-OES Service Manual
Agilent Restricted
135