HP 8712C User Manual page 398

Rf network analyzers
Table of Contents

Advertisement

Cable
and SRL measurements. See your Option 100 User's Guide Supplement for
information.
the current measurement mode.
See Chapter 6 for information on calibrating the analyzer.
Cal, mieck
check feature will compute and display corrected measurement uncertainties
(residual errors) that apply to the current instrument settings and
calibrations. This information may be used as an analysis or troubleshooting
tool when making transmission and reflection measurements.
See Chapter 6 for more information on using cal check.
Cal Kit
(default), Type-N male, 3.5 mm (for use with 50 fl systems), Type-F female
(for use with 75 fl systems), or user-defied.
By convention, cal kits indicate the sex of the pot% with which they are used.
For example the default cal kit for the analyzer is type-N
front panel RF ports are female (the calibration standards, in turn, are male).
This same convention applies to whatever test "port" the standard is attached
to, be it an adapter, cable, fixture, etc.
See Chapter 6 for more information on calibrating the analyzer.
measurement calibration along with the current instrument state.
See "Saving and Recalling Measurement Results" in Chapter 4 for information
on saving instrument states and measurement results to files.
Access Keys:
(
)
SAVE RECALL
Reference
C
because the
9-13