To Move the Pass/Fail
Indicator Text and Icon
To Turn the Pass/Fail
Indicator Text and Icon
measurement channel number and the word "pass" or "fail," such as 1: FAIL)
On and Off
by pressing (DISPLAY) Limit Mez~u Limit Options I&&% T&t ON off .
pressing @KKi$ Limit Fi%llLU Limit Options Limit Ican ON off . .
Using a Marker During
Pressing the C-1 softkey while entering a limit line will activate a
limit Entry
marker. The marker can be used to determine a frequency or level of interest
on the device response.
4-46
Using Instrument Functions
Using limit Testing
The limit test pass/fail indicator and text can be moved to any position on the
display screen.
lb move the position of the pass/fail indicator:
can also place the indicator along the horizontal by entering a percentage
using the numeric keypad. Enter a whole number from 0 to 100, where 0
represents the far left of the display, and 100 represents the far right of
the display.
Press I&it Raft y Position and use the front panel knob or the
can also place the indicator along the vertical by entering a percentage
using the numeric keypad. Enter a whole number from 0 to 100, where
0 represents the bottom of the display, and 100 represents the top of the
display.
You can toggle on or off the pass/fail indicator text (which contains the
You can toggle on or off the fail icon (there is no "pass" icon-only text) by