Automating Measurements
Automated Measurement Setup and Control
Recalling a state from RAM disk typically takes 4 seconds, but the time is
dependent on settings such as number of points.
One strategy for managing a large set of recall states is to initially store
them on a floppy disk. The files can then be copied from the floppy disk to
the volatile RAM disk at the start of each day. For example, to copy the file
For more details, see the Programmer's Guide.
When controlling the analyzer with a computer, you may want to copy a state
flle from your computer's disk to the analyzer's built-in RAM disk or floppy
disk. Later, the instrument can be instructed to recall the state from it's
internal disk. The lile can be sent from the computer to the analyzer's disk
using the SCPI MMEMory : TRANsf er commands. For more details, see the
Programmer's Guide.
For manually controlled test systems, the instrument state files for each
measurement can be stored onto disk by the test system designer. During the
can
(
)
test sequence, the operator
press the
key and recall the
SAVE RECALL
state corresponding to a particular measurement.