When reading the Cal l&tile, the analyzer recognizes only the information
after the "!" and ignores everything aftera "$". You can add comments after
a"$". Thetistlineofacalkithlemustcontaina"!" witha"$"chrectly
after it. No characters are allowed between the .!" and the "$" on the first
line ofacalkitfile.
10
!$
Standard Definitions for HP 85054B Precision Type-N Cal Kit.
20
!
30
!$
Definitions for 50 Ohm jack (FEMALE center contact) test
40
ports, plug (MALE center contact) standards.
!$
50
!
60
OPEN:
$ HP 85054-60027 Open Circuit Plug
!
z o 5 0 . 0 $ ohms
70
!
80
DELAY 57.993E-12 $ Set
!
90
LOSS 0.8E+9 $ Ohms/Set
!
100
CO 88.308E-15 $ Farads
!
110
!
Cl 1667.2E-27 $ Farads/Hz
120
!
130
!
140
!
150
!
SHORT:
zo 50.0 $ ohms
160
!
170
!
DELAY 63.078E-12 $ Set
LOSS 8.E+8 $ Ohms/Set
180
!
190
!
200
LOAD:
$ HP 00909-60011 Broadband Load Plug
!
210
zo 50.0 $ ohms
!
DELAY 0.0 $ Set
220
!
230
!
LOSS
240
!
250
THRU:
$ HP 85054-60038 Plug to Plug Adapter
!
260
zo 50.0 ohms
!
DELAY 196.OE-12 $ Set
270
!
LOSS
280
!
290
!
300
END
$ HP 85054-60025 Short Circuit Plug
0.0 $ Ohms/Set