HP 8712C User Manual page 143

Rf network analyzers
Table of Contents

Advertisement

To Use RF Filter
The RF Elter statistics function measures both the passband and the stopband
Statistics
(reject-band) of a hher with a single sweep.
3. At the end of each sweep, this feature calculates the insertion loss and
4. See Figure 4-14 for an example of an RF Elter stats search.
1. On measurement channel 1 press C-1 and place marker 1 at the
beginning of the passband and marker 2 at the end of the passband. Place
markers 3 and 4 at the beginning and end of the stopband. (When using
measurement channel 2, use markers 5 through 8 to deEne the passband
and stopband.)
peak-to-peak ripple of the passband, as well as the maximum signal
amplitude in the stopband. The insertion loss is deEned as the minimum
point between markers 1 and 2 with respect to 0 dB. The peak-to-peak
ripple of the passband is defined as the difference between the maximum
and minimum points in the passband (as deEned by markers 1 and 2). The
reject parameter is deEned as the difference between the minimum point
in the passband and the maximum point in the stopband.
Using Instrument Functions
Using Markers
4-25

Advertisement

Table of Contents
loading

This manual is also suitable for:

8714c

Table of Contents