Automating Measurements
Automated Measurement Setup and Control
Using learn Strings to
The IEEE 488 *LRN ("learn") command can also be used to set or query a
Save and Recall
complete set of instrument parameters. This can be used as a programming
Instrument States
convenience, eliminating the need for using disk files when saving and
recalling instrument states.
'lb obtain the learn string containing the instrument state, use the command
Since *LRN?, by IEEE definition, only contains the actual instrument state,
exclusive of data traces and calibration arrays, the network analyzer provides
the command
This command saves the data traces and calibration arrays if they are
enabled under
SCPI commands. Using the *LRN command to set the instrument settings
takes about the same amount of time as recalling a file from disk using -
MMEM:LOAD:STATe.
For more details on learn strings, refer to "Example Programs" in the
Programmer's Guide.
7-48
10 DIM Statel$ C40001
20 OUTPUT (PRfna; "*LRN?"
30 ENTER (PRfna USING "-K";Statel$
160 'I'Put the learn string back
170 OUTPUT ORfna;Statel$
SYST:SET:LRNLong?
(
]
SAVE RECALL
Define Saws or using the MMEM : STOR: STATE