7.8

Installing the Cassette Set Unit

1Y
NOTE
To install the cassette set unit, move the IP removal arm to the position shown in the illustration
below.
(1) Install the cassette set unit.
#2
[Attach] BR4x8 (x3)
Cassette set unit
009-058-00
10.20.2000 FM2732 (2)
At intervals of one year, or when 60,000 IPs have been processed
IP removal unit
CR-IR347 Service Manual
#1 [Rotate]
Timing belt
IP removal arm
PM - 40
Install.
Cassette
set unit
FR7H3315.EP
Positioning
screw
FR7H6100.EPS