Oscilloscope Setup - IBM 5500 Operational, Installation, And Maintenance Manual

Express ip control unit
Hide thumbs Also See for 5500:
Table of Contents

Advertisement

Oscilloscope Setup

116
Chapter 6: Twinax Client Cabling Requirements and Problem Solving
Figure 45: Measuring Distance Using an Oscilloscope
t
6 divisions × 0.2 µs = 1.2 µs
Df
1.2 µs × 324.7 ft = 390 ft
Dm
1.2 µs × 99 m = 119 m
The fault is indicated by the direction of the signal change (up for an open,
down for a short).
Note: If the station protectors are installed, the B-gate pulse may cause the
diodes to fire and generate a glitch on the displayed waveform.
Use an X1 probe (you can use X10 probes by changing the vertical input
setting) and the following extra parts:
• One coaxial cable with a BNC end and alligator clips (IBM P/N
1650790); or a BNC-to-banana plug adapter and multimeter leads
• One resistor equal to the impedance of the cable (110- Ohms resistor); or
resistor assembly (IBM P/N 7362344)
• One BNC T-connector (IBM P/N 1650789)
• One probe tip-to-BNC adapter (IBM P/N 453199).
Use Figure 47 on page 117 and do the following steps in order:
1. Connect the T-connector to the B-gate on the side panel of the
oscilloscope.
2. Use the probe tip adapter to connect the channel 1 probe to one side of the
T-connector.
3. Connect the cable to be tested on the other side of the T-connector. If
needed, use the coaxial cable with alligator clips or the BNC-to-banana
plug adapter with CE meter leads. Connect the two alligator clips to the
Planning, Installation, and Maintenance Guide
55E1I022

Advertisement

Table of Contents
loading

This manual is also suitable for:

5500-01e5500-02e5500-01t5500-02t

Table of Contents