Getting Started With X Series Usb Devices; Usb Device Chassis Ground - National Instruments DAQ X Series User Manual

Hide thumbs Also See for DAQ X Series:
Table of Contents

Advertisement

Getting Started with X Series USB Devices

The following sections contain information about X Series USB device
best practices and features.

USB Device Chassis Ground

Note (NI USB-636x Mass Termination Devices)
chassis ground connection through the I/O connector.
(NI USB-634x/635x/636x Screw Terminal Devices)
chassis of the X Series USB Screw Terminal device must be connected to
earth ground through the chassis ground.
The wire should be AWG 16 or larger solid copper wire with a maximum
length of 1.5 m (5 ft). Attach the wire to the earth ground of the facility's
power system. For more information about earth ground connections, refer
to the KnowledgeBase document, Earth Grounding for Test and
Measurement Devices, by going to
Code
You can attach and solder a wire to the chassis ground lug of the X Series
USB device, as shown in Figure 1-1. The wire should be as short as possible.
Figure 1-1. Grounding an NI USB-634x/635x/636x Screw Terminal Device through
© National Instruments
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
X Series USB Mass Termination devices have
.
earthground
the Chassis Ground Lug
1-3
Chapter 1
Getting Started
For EMC compliance, the
and entering the Info
ni.com/info
X Series User Manual