Chapter 4
Analog Input

Working Voltage Range

On most MIO X Series devices, the PGIA operates normally by amplifying
signals of interest while rejecting common-mode signals under the
following three conditions:
If any of these conditions are exceeded, the input voltage is clamped until
the fault condition is removed.

Analog Input Ground-Reference Settings

MIO X Series devices support the following analog input ground-reference
settings:
The AI ground-reference setting determines how you should connect your
AI signals to the MIO X Series device. Refer to the
Input Signals
Ground-reference settings are programmed on a per-channel basis. For
example, you might configure the device to scan 12 channels—four
differentially-configured channels and eight single-ended channels.
MIO X Series devices implement the different analog input
ground-reference settings by routing different signals to the NI-PGIA. The
NI-PGIA is a differential amplifier. That is, the NI-PGIA amplifies (or
attenuates) the difference in voltage between its two inputs. The NI-PGIA
X Series User Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The common-mode voltage (V
AI <0..x> GND from AI <0..x>–, must be less than ±10 V. This V
a constant for all range selections.
The signal voltage (V
), which is equivalent to subtracting AI <0..x>+
s
from AI <0..x>–, must be less than or equal to the range selection of
the given channel. If V
is greater than the range selected, the signal
s
clips and information are lost.
The total working voltage of the positive input, which is equivalent to
(V
+ V
), or subtracting AI GND from AI <0..x>+, must be less than
cm
s
±11 V.
Differential mode—In DIFF mode, the MIO X Series device
measures the difference in voltage between two AI signals.
Referenced single-ended mode—In RSE mode, the MIO X Series
device measures the voltage of an AI signal relative to AI GND.
Non-referenced single-ended mode—In NRSE mode, the MIO
X Series device measures the voltage of an AI signal relative to one of
the AI SENSE or AI SENSE 2 inputs.
section for more information.
4-4
), which is equivalent to subtracting
cm
Connecting Analog
is
cm
ni.com