National Instruments NI 6731 Calibration Procedure

National Instruments NI 6731 Calibration Procedure

Hide thumbs Also See for NI 6731:

Advertisement

Quick Links

NI 6711/6713/6731/6733
C
ALIBRATION
Introduction
Refer to
Note
ni671xCal.dll
What Is Calibration?
Why Should You Calibrate?
CVI™, LabVIEW™, National Instruments™, NI™, ni.com™, and NI-DAQ™ are trademarks of National Instruments
Corporation. Product and company names mentioned herein are trademarks or trade names of their respective
companies. For patents covering National Instruments products, refer to the appropriate location: Help»Patents in your
software, the patents.txt file on your CD, or ni.com/patents.
ni.com
© 2002 National Instruments Corp. All rights reserved.
P
ROCEDURE
This document contains step-by-step instructions for calibrating
National Instruments 6711/6713/6731/6733 for PCI/PXI/CompactPCI
analog output (AO) devices. Use this calibration procedure in conjunction
with the
ni671xCal.dll
for calibrating NI 6711/6713/6731/6733 devices.
ni.com/support/calibrat/mancal.htm
file.
Calibration consists of verifying the measurement accuracy of a device
and adjusting for any measurement error. Verification is measuring the
performance of the device and comparing these measurements to the
factory specifications. During calibration, you supply and read voltage
levels using external standards, then you adjust the module calibration
constants. The new calibration constants are stored in the EEPROM.
The calibration constants are loaded from memory as needed to adjust
for the error in the measurements taken by the device.
The accuracy of electronic components drifts with time and temperature,
which can affect measurement accuracy as the device ages. Calibration
restores these components to their specified accuracy and ensures that the
device still meets NI standards.
file, which contains specific functions required
for a copy of the
May 2002
370555A-01

Advertisement

Table of Contents
loading

Summary of Contents for National Instruments NI 6731

  • Page 1 Corporation. Product and company names mentioned herein are trademarks or trade names of their respective 370555A-01 companies. For patents covering National Instruments products, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your CD, or ni.com/patents.
  • Page 2: Internal Calibration

    How Often Should You Calibrate? The measurement requirements of your application determine how often the NI 6711/6713/6731/6733 must be calibrated to maintain accuracy. NI recommends that you perform a complete calibration at least once every year. You can shorten this interval to 90 days or six months based on the demands of your application.
  • Page 3: Test Equipment

    For this calibration procedure, you need NI-DAQ version 6.9.2 or earlier installed on the calibration computer. NI-DAQ, which configures and controls the NI 6711/6713/6731/6733, is available at ni.com/downloads © National Instruments Corporation NI 6711/6713/6731/6733 Calibration Procedure...
  • Page 4 NI-DAQ supports a number of programming languages, including LabVIEW, LabWindows/CVI, Microsoft Visual C++, Microsoft Visual Basic and Borland C++. When you install the driver, you only need to install support for the programming language that you intend to use. You also need copies of the , and ni671xCal.dll ni671xCal.lib...
  • Page 5 This step is described in the Adjusting the NI 6711/6713/6731/6733 section. Re-verify the performance to ensure that the NI 6711/6713/6731/6733 is operating within its specifications after adjustment. © National Instruments Corporation NI 6711/6713/6731/6733 Calibration Procedure...
  • Page 6 Verifying the Performance of the NI 6711/6713/6731/6733 Verification determines how well the device is meeting its specifications. The verification procedure is divided into the major functions of the device. Throughout the verification process, refer to the tables in the Specifications section to see if the device needs adjustment.
  • Page 7 100 ns. Call with paramID set to GPCTR_Change_Parameter ND_COUNT_2 and paramValue set to to configure the counter to output a pulse with an on time of 100 ns. © National Instruments Corporation NI 6711/6713/6731/6733 Calibration Procedure...
  • Page 8 Call with signal and source set to Select_Signal to route the counter signal to the GPCTR0_OUT ND_GPCTR0_OUTPUT pin on the device I/O connector. Call with action set to to start the GPCTR_Control ND_PROGRAM generation of the square wave. The device begins to generate a 5 MHz square wave when completes execution.
  • Page 9: Using The Tables

    Polarity refers to the positive and negative voltages of the input signal that can be read. Bipolar means the device can read both positive and negative voltages. Unipolar means that the device can read only positive voltages. © National Instruments Corporation NI 6711/6713/6731/6733 Calibration Procedure...
  • Page 10: Test Point

    Pos FS 9.9900000 9.9822988 9.9977012 9.9818792 9.9981208 Bipolar Neg FS –9.9900000 –9.9977012 –9.9822988 –9.9981208 –9.9818792 Table 4. Analog Output for the NI 6731/6733 Test Point 24-Hour Ranges 1-Year Ranges Lower Upper Lower Upper Range (V) Polarity Location Value (V) Limit (V)
  • Page 11 NI 6711/6713/6731/6733. Refer to the Calibrating the NI 6711/6713/6731/6733 section, the NI-DAQ Function Reference Help (Start»Programs»National Instruments»NI-DAQ» NI-DAQ Help), and the NI-DAQ User Manual for PC Compatibles at for additional information on the software structure. ni.com/manuals...
  • Page 12 Verifying the Counter GPCTR_Control (deviceNumber,0,ND_RESET) GPCTR_Set_Application (deviceNumber,0,ND_PULSE_TRAIN_GNR) GPCTR_Change_Parameter (deviceNumber,0,ND_COUNT_1,2) GPCTR_Change_Parameter (deviceNumber,0,ND_COUNT_2,2) Select_Signal (deviceNumber,ND_GPCTR0_OUTPUT,ND_GPCTR0_OUTPUT,ND_LOW_TO_HIGH) GPCTR_Control (deviceNumber,0,ND_PROGRAM) Figure 2. Verifying the Counter NI 6711/6713/6731/6733 Calibration Procedure ni.com...
  • Page 13 Adjusting the NI 6711/6713/6731/6733 Calibrate_E_Series (deviceNumber,ND_SELF_CALIBRATE,ND_USER_EEPROM_AREA,0) Get_Cal_Date (deviceNumber,*CalDate) Calibrate_E_Series (deviceNumber,ND_EXTERNAL_CALIBRATE,ND_USER_AREA,5.0) Copy_Const (deviceNumber) Figure 3. Adjusting the NI 6711/6713/6731/6733 © National Instruments Corporation NI 6711/6713/6731/6733 Calibration Procedure...

This manual is also suitable for:

Ni 6711Ni 6713Ni 6733