A programmable internal counter divides down the AO Sample Clock
Timebase signal.
Several other internal signals can be routed to AO Sample Clock through
internal routes. Refer to Device Routing in MAX in the NI-DAQmx Help or
the LabVIEW Help for more information.

Using an External Source

Use one of the following external signals as the source of AO Sample
Clock:
Routing AO Sample Clock Signal to an Output
Terminal
You can route AO Sample Clock (as an active low signal) out to any
PFI <0..15>, RTSI <0..7>, or PXIe-DSTARC terminal.

Other Timing Requirements

The AO timing engine on your device internally generates AO Sample
Clock unless you select some external source. AO Start Trigger starts the
timing engine and either the software or hardware can stop it once a finite
generation completes. When using the AO timing engine, you also can
specify a configurable delay from AO Start Trigger to the first AO Sample
Clock pulse. By default, this delay is two ticks of AO Sample Clock
Timebase.
© National Instruments
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
PFI <0..15>
RTSI <0..7>
PXI_STAR
PXIe-DSTAR<A,B>
Analog Comparison Event (an analog trigger)
5-11
Chapter 5
Analog Output
X Series User Manual