Chapter 7
Counters
Table 7-6 and Figure 7-34 detail a finite implicit generation of three
samples.

Continuous Buffered Implicit Pulse Train Generation

This function generates a continuous train of pulses with variable idle and
active times. Instead of generating a set number of data samples and
stopping, a continuous generation continues until you stop the operation.
Each point you write generates a single pulse. All points are generated back
to back to create a user defined pulse train.
Finite Buffered Sample Clocked Pulse Train
Generation
This function generates a predetermined number of pulse train updates.
Each point you write defines pulse specifications that are updated with each
sample clock. When a sample clock occurs, the current pulse (idle followed
by active) finishes generation and the next pulse updates with the next
sample specifications.
Note
When the last sample is generated, the pulse train continues to generate with these
specifications until the task is stopped.
X Series User Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Table 7-6. Finite Implicit Buffered Pulse Train Generation
Sample
Idle Ticks
1
2
3
Counter
Load Values
1 0 1 0 2 1 0 3
SOURCE
OUT
2
2
Counter Armed
Figure 7-34. Finite Implicit Buffered Pulse Train Generation
7-36
Active Ticks
2
2
3
4
2
2
2
1 0 1 0 1 0
3
4
2
2
ni.com

Hide quick links: