Chapter 6
Digital I/O

Using an Analog Source

When you use an analog trigger source, the acquisition begins on the first
rising or falling edge of the Analog Comparison Event signal.

Routing DI Start Trigger to an Output Terminal

You can route DI Start Trigger out to any PFI <0..15>, RTSI <0..7>, or
PXIe-DSTARC terminal. The output is an active high pulse. All PFI
terminals are configured as inputs by default.
The device also uses DI Start Trigger to initiate pretriggered DAQ
operations. In most pretriggered applications, a software trigger generates
DI Start Trigger. Refer to the DI Reference Trigger Signal section for a
complete description of the use of DI Start Trigger and DI Reference
Trigger in a pretriggered DAQ operation.

DI Reference Trigger Signal

Use the DI Reference Trigger (di/ReferenceTrigger) signal to stop a
measurement acquisition. To use a reference trigger, specify a buffer of
finite size and a number of pretrigger samples (samples that occur before
the reference trigger). The number of posttrigger samples (samples that
occur after the reference trigger) desired is the buffer size minus the
number of pretrigger samples.
Once the acquisition begins, the DAQ device writes samples to the buffer.
After the DAQ device captures the specified number of pretrigger samples,
the DAQ device begins to look for the reference trigger condition. If the
reference trigger condition occurs before the DAQ device captures the
specified number of pretrigger samples, the DAQ device ignores the
condition.
If the buffer becomes full, the DAQ device continuously discards the oldest
samples in the buffer to make space for the next sample. This data can be
accessed (with some limitations) before the DAQ device discards it. Refer
to the KnowledgeBase document, Can a Pretriggered Acquisition be
Continuous?, for more information. To access this KnowledgeBase, go to
ni.com/info
X Series User Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
and enter the Info Code
rdcanq
6-10
.
ni.com