SECTION 3: FUNCTIONAL BLOCK AND CIRCUIT ANALYSIS

INTRODUCTION
FUNCTIONAL BLOCK ANALYSIS
Oximetry Module
Introduction
Functional Block Analysis
Circuit Analysis
This section explains how monitor circuits operate and contains two parts.
Functional block diagram analysis
Circuit analysis
Monitor functional elements (figure FOG-1) include:
Standard original equipment manufacturer (OEM) oximetry module
(SOOM), model MP-203 oximetry module. For ease of reference this
module is hereafter referred to as the oximetry module. In actual fact,
the oximetry module is a printed circuit board (PCB).
CPU PCB
LED Drive PCB
LED PCB
Front panel controls
power inlet
AC
Battery
The oximetry module (figure 3-1) drives the oxygen transducer and conditions
the signal derived from the patient (via the sensor connector).