Technical Supplement
S2.3
Measured versus Calculated Saturation
S3
CIRCUIT ANALYSIS
S-2
When saturation is calculated from a blood gas measurement of the partial
pressure of arterial oxygen (PaO
NPB-40 SpO
measurement. This is because the calculated saturation may
2
not have been corrected for the effects of variables that can shift the
relationship between PaO
Figure S2-1 illustrates the effect that variations in pH, temperature, partial
pressure of carbon dioxide (PCO
hemoglobin may have on the oxyhemoglobin dissociation curve.
100
Fetal Hb
50
0
Figure S2-1: Oxyhemoglobin Dissociation Curve
This section provides an explanation of NPB-40 theory of operation using
block diagrams and schematic diagrams.
The NPB-40 consists of three main functional components described in the
following paragraphs:
The CPU PCB block diagram (Figure S3-2) and schematic diagram
(Figure S4-3).
The LCD PCB block diagram (Figure S3-3) and schematic diagram
(Figure S4-4).
The Front Panel PCB schematic diagram (Figure S4-5).
The relationship between these components and their interconnection is
illustrated in the NPB-40 block diagram (Figure S3-1).
), the calculated value may differ from the
2
and saturation.
2
), and concentrations of 2,3-DPG and fetal
2
pH
Temperature
PCO
2
2,3-DPG
50
PO
(mmHg)
2
pH
Temperature
PCO
2
2,3-DPG
100