Characteristics and Specifications
358

Trigger system

Sources
Modes
Holdoff time range
Trigger jitter
Selections
Scope channel triggering
Range (internal)
Sensitivity*
Coupling
* Denotes warranted specifications, all others are typical. Specifications are valid
after a 30-minute warm-up period and ±10 ° C from firmware calibration
temperature.
Logic (D15 - D0) channel triggering (with MSO option)
Threshold range (user defined) ±8.0 V in 10 mV increments
Threshold accuracy*
Predefined thresholds
* Denotes warranted specifications, all others are typical. Specifications are valid
after a 30-minute warm-up period and ±10 ° C from firmware calibration
temperature.
External (EXT) triggering
Input resistance
Maximum input
Sensitivity
Coupling
Probe ID
DSO6xx4L: Ch 1, 2, 3, 4, line, ext and D0 - D15 for
MSO enabled DSO
Auto, Normal, Single
~60 ns to 10 seconds
15 psec rms
Edge, pulse width, pattern, TV, duration,
sequence, CAN, LIN, USB, I2C, SPI, Nth edge
burst
±6 div from center screen
< 10 mV/div: greater of 1 div or 5 mV
≥ 10mV/div: 0.6 div
AC (~10 Hz), DC, noise reject, HF reject and LF
reject (~ 50 kHz)
±(100 mV + 3% of threshold setting)
TTL = 1.4 V, CMOS = 2.5 V, ECL = -1.3 V
1.015 kΩ ± 5% (DSO6014L)
2.14 kΩ ± 5% (DSO6054L/6104L)
±15 V
Range ±5 V
DC to 100 MHz: 500 mV (DSO6014L)
DC to 500 MHz: 500 mV (DSO6054L/6104L)
AC (~ 3.5 Hz), DC, noise reject, HF reject and LF
reject (~ 50 kHz)
Auto probe sense (DSO6014L)
Auto probe sense and AutoProbe interface
(DSO6054L/6104L)
6000 Series Oscilloscope User's Guide