Binary — Binary logic levels are displayed directly above
the softkeys for the current X1 and X2 cursor positions
for all displayed channels.
Cursor X1 values
Cursor X1 values for
for analog channels
digital channels
1 through 4
D15 Through D0
Hex — Hexadecimal logic levels are displayed directly
above the softkeys for the current X1 and X2 cursor
positions for all displayed channels.
In hexadecimal and binary mode, a level can display as 1
(higher than trigger level), 0 (lower than trigger level),
indeterminate state ( ), or X (don't care). In binary
mode, X is displayed if the channel is turned off. In hex
mode, the channel is interpreted as a 0 if turned off.
3 Press the Source softkey to select the analog channel or
math source on which the Y cursors will indicate
measurements.
The source in Normal cursor mode can be any analog
channel or math source. If you select binary or
hexadecimal mode, the Source softkey is disabled since
you are displaying binary or hexadecimal levels for all
channels.