6000 Series Oscilloscope User's Guide
The
oscilloscope's MegaZoom III technology gives you bright,
easily viewed displays of any part of the video waveform.
Analysis of video waveforms is simplified by the oscilloscope's
ability to trigger on any selected line of the video signal.
1 Press the More key in the Trigger section of the front panel. If
TV is not selected, rotate the Entry knob until TV is displayed
in the Trigger softkey, then press the Settings softkey to
display TV trigger menu.
Source
Sync
channel
polarity
2 Press the Source softkey and select any analog channel as the
TV trigger source.
Triggering the Oscilloscope
TV
trigger
TV Standard
TV mode
Selected
channel
TV Mode
Line #
Return to
previous menu