4
Triggering the Oscilloscope
154
4 Press the BUSDOCTOR 2 softkey to access the BUSDOCTOR 2
Menu.
Mode
Baud Rate
Specifying the BUSDOCTOR LAN Address
This is necessary when the oscilloscope controls BUSDOCTOR
(but not when a PC with VISION software controls
BUSDOCTOR).
1 From the BUSDOCTOR 2 Menu, press the LAN Address softkey
to access the BUSDOCTOR LAN Addresses Menu.
2 Enter the BUSDOCTOR's 4-byte IP address:
a Press the Value softkey and use the Entry knob to select a
byte value.
b Press the Next softkey to access the next byte.
3 When you have finished entering the IP address, press the
Apply softkey.
This is sufficient for connection via a hub or switch.
If the oscilloscope is connected directly to the BUSDOCTOR (in
a point-to-point configuration), the following additional steps
must be taken.
TDMA Struct
Channel
(Sync Mode)
6000 Series Oscilloscope User's Guide
FlexRay
trigger
BUSDOCTOR
Return to
LAN Address
previous menu